Technical Expertise + Quality Products + Lowest Prices = The Best Solution
Thermco Gas Analyzer Model: 6900
This item is custom built to fit your needs and has various part numbers. Please contact us for pricing and...
This item is custom built to fit your needs and has various part numbers. Please contact us for pricing and expected ship date.
Model | 6900 |
Ranges |
0-100% CO2 in Ar
0-100% O2 in Ar 0-100% He in Ar 0-100% H2 in Ar 0-100% N2 in Ar 0-100% CO2 in N2 0-100% O2 in N2 0-100% He in N2 0-100% H2 in N2 |
Accuracy |
Valid over 50°F to 90°F (10°C to 32°C) ambient temperature conditions.
± 0.3% of minor component over range of 0-10.0%, all ranges. Example: for a 5% O2 in Ar mixture, the analysis will be between 4.7 and 5.3% O2. ±0.5% of minor component over a range of 10.1% to 30.0%, all ranges. Example: for a 25% CO2 in Ar mixture, the analysis will be between 24.5% and 25.5% CO2. ±1% of minor component over a range of 30.1% to 60.0%, all ranges. Example: for a 50% CO2 in N2 mixture, the analysis will be between 49.0% and 51.0% CO2. ±2% of minor component over a range of 60.1% to 100.0%, all ranges, except CO2 in Ar, which has an accuracy of ±8%. Example: for a 75% He in Ar mixture, the analysis will be between 73.0% and 77.0% He. |
Operating Temperatures | 40°F to 100°F (4°C to 38°C) |
Resolution | 0.1% |
Response Time | 95% in 45 seconds for typical gas mixtures |
Utilities Required | Reference Gas: Commercial grade argon or nitrogen (depending upon the base gas in range used) is required to flow continuously at 0.5 SCFH during analysis. Sample Gas: Moisture and oil free sample gas is required to flow continuously at 1.0 SCFH during analysis. Sample, reference gas pressure: 1-100 PSIG Power: 115 VAC, 60 Hz., 1 Amp Optional – Model 6900-100, 100 VAC, 50/60 Hz, 1 ampere Optional – Model 6900-230, 230 VAC, 50/60 Hz, 0.5 ampere |
Frequency of Calibration | Calibration is recommended once per year when used under spot checking conditions; if run continuously will depend on conditions and range. |
Warm Up Time | 8 minutes under room temperature conditions |
Dimensions and Weight | 9" High, 11" Wide, 14" Deep (23 cm x 28 cm x 35.5 cm) |
Weight | 22 lbs. (10 kg) |
Dimensions | 9.0 in (23.0 cm) height, 11.0 in (28.0 cm) width, 14.0 in (35.5 cm) depth |
Altitude | Sea level to 2000 ft. (Altitude above 2000 ft. will cause some inaccuracy; consult the factory) |
Analyzer Outputs | Current: 4-20 mA, proportional to gas analysis range selected (available on all ranges). The 4-20 mA resolution equals display resolution of 0.1% gas analysis. |
Additional Information
Custom Range
The user may encounter a gas mixture pair other than the nine standard ranges built into the analyzer. In this case, the user may perform an analysis by inputting his own calibration data into the analyzer. This can be accomplished by utilizing the pushbuttons on the front meter. Guidance is available from the factory concerning the thermal conductivity response of a wide variety of gas pairs.
Service
Thermco has been providing service for their analyzers since 1951. The Thermco repair department can normally complete repairs in two weeks.
Warning
Improper use of this product can cause serious injury or property damage. Personnel dealing with this equipment should read and understand warning labels and instruction manuals provided by Thermco.
The Model 6900 is not intended to measure oxygen/nitrogen mixtures for breathing air, oxygen purity, or other medical applications.
The Model 6900 is not designed to be used with toxic gases or gases corrosive to brass or nylon.